Black & Decker MULTI-GAS DETECTORS User Manual Page 47

  • Download
  • Add to my manuals
  • Print
  • Page
    / 72
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 46
GasAlertMax
Calibration and Setting Alarm Setpoints
Auto Span
The LCD displays K prompting you to apply the calibration
gas to the sensors or you can skip the span (sensitivity
adjustment).
Use a calibration gas meeting the recommended gas mixture
as listed in the guidelines.
5. Apply gas to the sensor at a flow rate of 250 to
500 ml/min.
OR
Press C to skip the span.
When the detector senses approximately one-half of the
expected gas concentration (30 seconds), the audible alarm
beeps once. The detector then begins spanning the sensor
(2 minutes). (These times are valid if a GasAlertMax
calibration gas is used.)
The audible alarm beeps three times at the end of the span.
Note
The detector will not span a sensor if:
You do not apply gas to the sensor.
The sensor fails to detect at least one-half of
the expected gas concentration in the first 30
seconds.
The gas concentration drops below one-half of
the expected gas level during the 2-minute
span.
If you apply gas to a sensor and the detector fails to
span the sensor, press
C
to exit auto span. Then
press
F
to exit the calibration and repeat the
calibration process using a new gas cylinder. If the
sensor fails the span a second time, replace the
sensor. Refer to the section,
Replacing a Sensor,
Pump, or Pump Filter.
37
Page view 46
1 2 ... 42 43 44 45 46 47 48 49 50 51 52 ... 71 72

Comments to this Manuals

No comments