SAMPLING PUMP
1. Attach the Sampling Hose or Sampling Probe directly to the Sampling Pump. Turn
the quick connect right to lock into place.
TO ACTIVATE THE PUMP
2. Simply slide the thumbswitch fully to the
right closing the vents. The pump will
automatically turn ON, drawing the
sampled air across the sensors. The
Sampling Pump Fan Icon will light.
During pump operation, the BW
DEFENDER will display the FAN icons in
a “spin” pattern. This indicates that the
pump is operating normally.
Allow 2 minutes or more for the sample to
travel up the sample tubing (dependent on
the length of the tubing).
TO RETURN TO AMBIENT MODE
3. Slide the thumbswitch fully left, opening the vents, which allows the ambient air to
reach the sensors via normal diffusion. The pump will automatically turn OFF.
LOW FLOW ALARM
GENERAL: Should the Hose (Probe) or Pump become clogged or plugged, the
BW DEFENDER unit will go into alarm. All Alarm indicators will pulse rapidly.
VISUAL ALARMS: The FAN icon will flash ON and OFF
The “HIGH ALARM” icon will flash
The LED Alarms will flash
AUDIBLE ALARMS: The Audible Beeper will sound in a fast intermittent
pulse
1. Turn the pump OFF immediately. Slide the Pump's thumbswitch fully left opening
the vents.
2. Clear obstruction before proceeding. Clean out clogged tubing probe by blowing it
out or flushing with clean water and drying. Wash and dry clogged blue Particulate
Filter (or replace). Check the clear cap to see if water has reached the pump trap
(see page 18).
SAMPLING PUMP
HYDROPHOBIC FILTER: The replaceable hydrophobic Teflon filter should be
inspected periodically to ensure that it is clean. Dirty filters will impede airflow into the
BW DEFENDER unit, and should be replaced as soon as possible.
WATER INVASION: Due to design, water
is unlikely to reach the pump housing. But,
in the event it does, open the housing and
drain the liquid from the built-in water trap.
Dry the filter.
1. Remove the 3 screws holding the clear
cap in place.
2. Drain out any liquid and air dry the filter.
If the filter is dirty or damaged, remove
the old filter.
3. Put the new filter in place overtop the
adhesive ring and lightly press into
place. Be certain the filter does not
obscure the nutserts and is correctly
aligned.
4. Replace the clear cap, ensuring the
three Phillips screws are fully seated
and the filter is in the correct position.
Hydrophobic Filter Kit of 10 (Order #: M1864K)
SAMPLING PROBES
The 1, 3 or 6 ft. (0.3, 0.9 or 1.8 m)
Probes are equipped with a handle and
quick connect.
SAMPLE HOSE
D4-SP01 will include a D4-HSP-10, 10 ft.
(3 m) sampling hose (longer lengths
available). The Sampling Hose is
equipped with a blue Particulate Filter
(Order #: D4-PF-1) and a quick connect.
Sample hoses are available up to
15.24 m (50 ft.) in length.
17 18